天天摸天天碰天天添中文字幕,国产区点击进入在线影院,无码人妻一区二区免费AV,亚洲精品国产av现线

產(chǎn)品中心

溶血磷脂酰膽堿(LPC)酶聯(lián)免疫吸附檢測試劑盒
ELK8145
規(guī)格: 價(jià)格:
48T ¥2240.00
96T ¥3200.00

Overview 文獻(xiàn)

Product name: LPC(Lysophosphatidyl choline) ELISA Kit
Reactivity: General
Alternative Names: lysoPC; Lysolecithin; 1-Palmitoyl-sn-Glycero-3-Phosphocholine
Assay Type: Competitive Inhibition
Sensitivity: 92.4 ng/mL
Standard: 20000 ng/mL
Detection Range: 312.5-20000 ng/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 2h
Research Area: Enzyme & Kinase;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with LPC. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to LPC. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of LPC in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (ng/mL) OD Corrected OD
20000.00 0.212
10000.00 0.357
5000.00 0.611
2500.00 0.823
1250.00 1.157
625.00 1.511
312.50 1.796
0.00 2.112

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant LPC and the recovery rates were calculated by comparing the measured value to the expected amount of LPC in samples.
Matrix Recovery range Average
serum(n=5) 91-105% 98%
EDTA plasma(n=5) 85-97% 91%
Heparin plasma(n=5) 88-101% 88%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of LPC and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 88-102% 86-92% 83-98% 96-103%
EDTA plasma(n=5) 85-94% 97-103% 82-90% 92-101%
Heparin plasma(n=5) 85-97% 86-102% 87-103% 85-101%